Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Tests and Calibration of the NIF Neutron Time of Flight Detectors

Conference ·
OSTI ID:939696

The National Ignition Facility (NIF) Neutron Time of Flight (NTOF) diagnostic will measure neutron yield and ion temperature in all NIF campaigns in DD, DT, and THD* implosions. The NIF NTOF diagnostic is designed to measure neutron yield from 109 to 2 x 1019. The NTOF consists of several detectors of varying sensitivity located on the NIF at about 5 m and 20 m from the target. Production, testing, and calibration of the NIF NTOF detectors have begun at the Laboratory for Laser Energetics (LLE). Operational tests of the NTOF detectors were performed on several facilities including the OMEGA laser at LLE and the Titan laser at Lawrence Livermore National Laboratory (LLNL). Neutron calibrations were carried out on the OMEGA laser. Results of the NTOF detectors tests and calibration will be presented. *(D = deuterium, T = tritium, H = hydrogen)

Research Organization:
National Security Technologies, LLC
Sponsoring Organization:
USDOE National Nuclear Security Administration (NA)
DOE Contract Number:
AC52-06NA25946
OSTI ID:
939696
Report Number(s):
DOE/NV/25946--451
Country of Publication:
United States
Language:
English

Similar Records

Tests and calibration of NIF neutron time of flight detectors
Journal Article · Wed Oct 15 00:00:00 EDT 2008 · Review of Scientific Instruments · OSTI ID:21266516

Tests and Calibration of NIF Neutron Time of Flight Detectors
Journal Article · Wed Nov 12 23:00:00 EST 2008 · REVIEW OF SCIENTIFIC INSTRUMENTS 79 · OSTI ID:940920

The NIF 4.5-m nTOF Detectors
Conference · Mon May 07 00:00:00 EDT 2012 · OSTI ID:1043655