Design considerations for a time-resolved tomographic diagnostic at DARHT
Conference
·
· Proceedings of SPIE, the International Society for Optical Engineering
An instrument has been developed to acquire time-resolved tomographic data from the electron beam at the DARHT [Dual-Axis Radiographic Hydrodynamic Test] facility at Los Alamos National Laboratory. The instrument contains four optical lines of sight that view a single tilted object. The lens design optically integrates along one optical axis for each line of sight. These images are relayed via fiber optic arrays to streak cameras, and the recorded streaks are used to reconstruct the original two-dimensional data. Installation of this instrument into the facility requires automation of both the optomechanical adjustments and calibration of the instrument in a constrained space. Additional design considerations include compound tilts on the object and image planes.
- Research Organization:
- National Security Technologies, LLC
- Sponsoring Organization:
- USDOE - National Nuclear Security Administration (NNSA)
- DOE Contract Number:
- AC52-06NA25946
- OSTI ID:
- 935203
- Report Number(s):
- DOE/NV/25946--009; DOE/NV/11718--1249
- Conference Information:
- Journal Name: Proceedings of SPIE, the International Society for Optical Engineering Journal Volume: 6289
- Country of Publication:
- United States
- Language:
- English
Similar Records
Data and Analysis from a Time-Resolved Tomographic Optical Beam Diagnostic
Fidelity of a Time-Resolved Imaging Diagnostic for Electron Beam Profiles
Fielding of a time-resolved tomographic diagnostic
Conference
·
Sun May 02 00:00:00 EDT 2010
·
OSTI ID:993425
Fidelity of a Time-Resolved Imaging Diagnostic for Electron Beam Profiles
Journal Article
·
Wed Oct 01 00:00:00 EDT 2014
· IEEE Transactions on Plasma Science
·
OSTI ID:1178117
Fielding of a time-resolved tomographic diagnostic
Conference
·
Fri Feb 29 23:00:00 EST 2008
·
OSTI ID:926231