- ORNL
A new technique is presented for the measurement of neutron and/or gamma-ray coincidences. Separate neutron neutron, neutron gamma-ray, gamma-ray neutron, and gamma-ray gamma-ray coincidences are acquired with liquid scintillation detectors and a digital pulse shape discrimination (PSD) technique based on standard charge integration method. The measurement technique allows for the collection of fast coincidences in a time window of the order of a few tens of nanoseconds between the coincident particles. The PSD allows for the acquisition of the coincidences in all particle combinations. The measurements are compared to results obtained with the MCNP-PoliMi code, which simulates neutron and gamma-ray coincidences from from a source on an event-by-event basis. This comparison leads to good qualitative agreement.
- Research Organization:
- Oak Ridge National Laboratory (ORNL)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC05-00OR22725
- OSTI ID:
- 932640
- Report Number(s):
- ORNL/TM-2007/196
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance of coincidence-based PSD on LiF/ZnS Detectors for Multiplicity Counting
Study of the filter method for neutron pulse-height distributions measured with organic scintillators
Real-time, digital pulse-shape discrimination in non-hazardous fast liquid scintillation detectors: Prospects for safety and security
Conference
·
Thu Oct 06 00:00:00 EDT 2016
·
OSTI ID:1334884
Study of the filter method for neutron pulse-height distributions measured with organic scintillators
Journal Article
·
Thu Jan 07 23:00:00 EST 2016
· Radiation Measurements
·
OSTI ID:1366591
Real-time, digital pulse-shape discrimination in non-hazardous fast liquid scintillation detectors: Prospects for safety and security
Conference
·
Fri Jul 01 00:00:00 EDT 2011
·
OSTI ID:22039844