TGLD: A 16-Channel Charge Readout Chip for the PHENIX Pad Chamber Detector Subsystem at RHIC

Conference ·
OSTI ID:931808
This paper describes TGLD, a charge readout chip for the PHENIX Pad Chamber (PC) subsystem at Brookhaven National Laboratory's Relativistic Heavy Ion Collider (RHIC) in Upton, NY. Due to the PC's high channel density, the TGLD and associated circuitry operate within the active detector region as permanent, zero access components, with remote set-up and test during collider operation. The TGLD design accommodates varying pad capacitance and charge gain for three detector subassemblies that detect particles at three different distances from the PHENIX collision vertex. The design also provides adjustable discrimination thresholds from MIP/10 to 2 MIP (Minimum Ionizing Particle). Three TGLD chips operate with a complimentary digital memory unit (DMU) to form 48 channel low power, low mass, readout cards. Partitioning of readout electronics and address control for robust remote operation are discussed. Component and system test results are also reported.
Research Organization:
Oak Ridge National Laboratory (ORNL)
Sponsoring Organization:
SC USDOE - Office of Science (SC)
DOE Contract Number:
AC05-00OR22725
OSTI ID:
931808
Country of Publication:
United States
Language:
English

Similar Records

TGLD: A 16-channel charge readout chip for the PHENIX Pad Chamber detector subsystem at RHIC
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:624216

FPHX: A New Silicon Strip Readout Chip for the PHENIX Experiment at RHIC
Conference · Wed Dec 31 23:00:00 EST 2008 · Submitted to IEEE Trans.Nucl.Sci. · OSTI ID:970995

A front-end electronics module for the PHENIX pad chamber
Journal Article · Tue Nov 30 23:00:00 EST 1999 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers) · OSTI ID:20015427