Urine is universally recognized as one of the best non-invasive matrices for biomonitoring exposure to a broad range of xenobiotics including toxic metals. For direct, simple, and field-deployable monitoring of urinary Pb, electrochemical sensors employing superparamagnetic iron oxide (Fe3O4) nanoparticles with a surface functionalization of dimercaptosuccinic acid (DMSA) has been developed. The metal detection involves rapid collection of dispersed metal-bound nanoparticles from a sample solution at a magnetic or electromagnetic electrode, followed by the stripping voltammetry of the metal in acidic medium. The sensors were evaluated as a function of solution pH, the binding affinity of Pb to DMSA-Fe3O4, the ratio of nanoparticles per sample volume, preconcentration time, and Pb concentrations. The effect of binding competitions between the DMSA-Fe3O4 and urine constituents for Pb on the sensor responses was studied. After 90s of preconcentration in samples containing 25 vol.% of rat urine and 0.1 g/L of DMSA-Fe3O4, the sensor could detect background level of Pb (< 1 ppb) and yielded linear responses from 0 to 50 ppb of Pb, excellent reproducibility (%R.S.D of 5.3 for seven measurements of 30 ppb Pb), and Pb concentrations comparable to those measured by ICP-MS. The sensor could also simultaneously detect background levels (< 1 ppb) of Cd, Pb, Cu, and Ag in river and seawater.

Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (US), Environmental Molecular Sciences Laboratory (EMSL)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
926934
Report Number(s):
PNNL-SA-57442; 24801; 400412000
Journal Information:
The Analyst, 133(3):348-355, Journal Name: The Analyst, 133(3):348-355
Country of Publication:
United States
Language:
English

Similar Records

Microanalyzer for Biomonitoring of Lead (Pb) in Blood and Urine
Journal Article · Sun Dec 31 23:00:00 EST 2006 · Analytical and Bioanalytical Chemistry, 387(1):335-341 · OSTI ID:898097

Detection of Cd, Pb, and Cu in non-pretreated natural waters and urine with thiol functionalized mesoporous silica and Nafion composite electrodes
Journal Article · Wed May 21 00:00:00 EDT 2008 · Analytica Chimica Acta, (620):55-63 · OSTI ID:935064

Renal handling of technetium-99m DMSA: Evidence for glomerular filtration and peritubular uptake
Journal Article · Sat Jul 01 00:00:00 EDT 1989 · J. Nucl. Med.; (United States) · OSTI ID:5601713