Fabrication and In Vitro Deployment of a Laser-Activated Shape Memory Polymer Vascular Stent

Journal Article · · BioMedical Engineering OnLine, vol. 6, n/a, November 27, 2007, pp. 43

Vascular stents are small tubular scaffolds used in the treatment of arterial stenosis (narrowing of the vessel). Most vascular stents are metallic and are deployed either by balloon expansion or by self-expansion. A shape memory polymer (SMP) stent may enhance flexibility, compliance, and drug elution compared to its current metallic counterparts. The purpose of this study was to describe the fabrication of a laser-activated SMP stent and demonstrate photothermal expansion of the stent in an in vitro artery model. A novel SMP stent was fabricated from thermoplastic polyurethane. A solid SMP tube formed by dip coating a stainless steel pin was laser-etched to create the mesh pattern of the finished stent. The stent was crimped over a fiber-optic cylindrical light diffuser coupled to an infrared diode laser. Photothermal actuation of the stent was performed in a water-filled mock artery. At a physiological flow rate, the stent did not fully expand at the maximum laser power (8.6 W) due to convective cooling. However, under zero flow, simulating the technique of endovascular flow occlusion, complete laser actuation was achieved in the mock artery at a laser power of {approx}8 W. We have shown the design and fabrication of an SMP stent and a means of light delivery for photothermal actuation. Though further studies are required to optimize the device and assess thermal tissue damage, photothermal actuation of the SMP stent was demonstrated.

Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
926014
Report Number(s):
UCRL-JRNL-230412
Journal Information:
BioMedical Engineering OnLine, vol. 6, n/a, November 27, 2007, pp. 43, Journal Name: BioMedical Engineering OnLine, vol. 6, n/a, November 27, 2007, pp. 43 Vol. 6
Country of Publication:
United States
Language:
English

References (38)

Optical properties of the clearest natural waters (200–800 nm) journal January 1981
Remotely actuated polymer nanocomposites—stress-recovery of carbon-nanotube-filled thermoplastic elastomers journal January 2004
Blood Viscosity and Elevated Carotid Intima-Media Thickness in Men and Women: The Edinburgh Artery Study journal April 1998
Photothermal properties of shape memory polymer micro-actuators for treating stroke journal January 2002
Overview of Bioheat Transfer book January 1995
Laser-activated shape memory polymer intravascular thrombectomy device journal January 2005
Thermal-perfusion balloon coronary angioplasty: In vivo evaluation journal January 1993
Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry journal January 1991
Merci ® Retriever in acute stroke treatment journal May 2006
Drug-Eluting Bioresorbable Stents for Various Applications journal August 2006
Drug-Eluting Stents: The Price Is Not Right journal October 2006
Shape Memory Polymer Stent With Expandable Foam: A New Concept for Endovascular Embolization of Fusiform Aneurysms journal June 2007
Prototype Fabrication and Preliminary In Vitro Testing of a Shape Memory Endovascular Thrombectomy Device journal September 2007
Compliance Matching Stent Placement in the Carotid Artery of the Swine Promotes Optimal Blood Flow and Attenuates Restenosis journal October 2004
Cold hibernated elastic memory foams for endovascular interventions journal February 2003
Correlation of HSP70 Expression and Cell Viability Following Thermal Stimulation of Bovine Aortic Endothelial Cells journal May 2005
Role of stent design and coatings on restenosis and thrombosis journal June 2006
Diffusion model of the optical absorbance of whole blood journal January 1988
Polyurethane Shape-Memory Polymers Demonstrate Functional Biocompatibility In Vitro journal January 2007
State of the art in coronary intervention journal February 2003
Unconstrained recovery characterization of shape-memory polymer networks for cardiovascular applications journal May 2007
Angioplasty and stenting of intracranial stenosis journal January 2005
Shape memory polymers based on uniform aliphatic urethane networks journal January 2007
Thermomechanics of the shape memory effect in polymers for biomedical applications journal June 2005
Long-term outcomes and predictors of iliac angioplasty with selective stenting journal September 2005
Volume and density changes of biological fluids with temperature journal December 1985
New Frontiers in Cardiology: Drug-Eluting Stents: Part I journal May 2003
Stent Struts and Articulations: Their Impact on Balloon-Expandable Stents’ Hoop Strength, Pushability, and Radiopacity in an Experimental Setting journal January 2002
Normal values of volume flow in the internal carotid artery measured by a new angle-independant Doppler technique for evaluating cerebral perfusion journal July 2004
Structure and properties of shape-memory polyurethane block copolymers journal May 1996
Inductively Heated Shape Memory Polymer for the Magnetic Actuation of Medical Devices journal October 2006
Collapse pressures of biodegradable stents journal May 2003
Rate process model for arterial tissue thermal damage: Implications on vessel photocoagulation journal January 1994
Intracranial angioplasty and stenting for cerebral atherosclerosis: new treatments for stroke are needed! journal April 2006
Early and late arterial healing response to catheter-Induced laser, thermal, and mechanical wall damage in the rabbit journal January 1990
Absence of biological damage from prolonged exposure to intravascular ultrasound: A swine model journal March 2007
Laser-activated shape memory polymer microactuator for thrombus removal following ischemic stroke: preliminary in vitro analysis journal July 2005
Shape-memory behavior of thermally stimulated polyurethane for medical applications journal January 2006

Similar Records

Endovascular Stent-Graft Applications in Iatrogenic Vascular Injuries
Journal Article · Mon Sep 15 00:00:00 EDT 2003 · Cardiovascular and Interventional Radiology · OSTI ID:21088357

Subintimal Double-Barrel Restenting of an Occluded Primary Stented Superficial Femoral Artery
Journal Article · Fri Jun 15 00:00:00 EDT 2007 · Cardiovascular and Interventional Radiology · OSTI ID:21090894

Carotid Artery Stenting Using a Novel Self-Expanding Braided Nickel-Titanium Stent: Feasibility and Safety Porcine Trial
Journal Article · Tue Sep 15 00:00:00 EDT 2009 · Cardiovascular and Interventional Radiology · OSTI ID:21429114