MGA++ Analysis of Low Quantity Samples of U and Pu on an Extended-rage Gamma-ray Detector
The IAEA has expressed a need for improved determination of gamma emitting nuclides in environmental samples collected during inspections of nuclear facilities and to use the MGA++ to determine U and Pu concentrations and isotopic compositions when those elements are present in relatively high concentrations. We are addressing the IAEA needs by evaluating the applicability of extended-range germanium detectors (ERG). In this paper we used 1g U isotopic standards and 100ug Pu liquid standards (1) to determine the performance of MGA++ on this special detector and (2) to estimate the amount of U and Pu necessary in the sample for determination of the isotopics via MGA++ within reasonable accuracy for a week of counting time using this ERG detector.