Title: Performance, Thermal, and Vibration Qualification Testing of Zetec Acoustic Transducers, Model Z0002659-2, Sondicator Probes

Abstract

This report is a result of Qualification Test Plan No.001 prepared by Anthony Lavietes. The Qualification Test Plan outlines a list of requirements for thermal and vibrational testing of Zetac's Z0002659-2 Sondicator Probe acoustic transducers (hereafter called ''transducers''). The Zetec transducers are used in a system that employs an array of 7 acoustic transducers. Qualification testing of these transducers was required since they are a modified version of a standard catalog item from the manufacturer. This report documents the thermal, vibrational, and performance testing that was performed on a sampling of these transducers in order to qualify them for flight. A total of 14 transducers were tested. All 14 passed qualification testing with no failures.

Authors:
; ;
Publication Date:
Research Org.:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Org.:
USDOE
OSTI Identifier:
888621
Report Number(s):
UCRL-TR-219738
TRN: US200618%%446
DOE Contract Number:  
W-7405-ENG-48
Resource Type:
Technical Report
Country of Publication:
United States
Language:
English
Subject:
47 OTHER INSTRUMENTATION; 42 ENGINEERING; ACOUSTICS; PERFORMANCE; PERFORMANCE TESTING; PROBES; SAMPLING; TESTING; TRANSDUCERS

Citation Formats

Jacobson, G, Gemberling, S, and Lavietes, A. Performance, Thermal, and Vibration Qualification Testing of Zetec Acoustic Transducers, Model Z0002659-2, Sondicator Probes. United States: N. p., 2006. Web. doi:10.2172/888621.
Jacobson, G, Gemberling, S, & Lavietes, A. Performance, Thermal, and Vibration Qualification Testing of Zetec Acoustic Transducers, Model Z0002659-2, Sondicator Probes. United States. doi:10.2172/888621.
Jacobson, G, Gemberling, S, and Lavietes, A. Fri . "Performance, Thermal, and Vibration Qualification Testing of Zetec Acoustic Transducers, Model Z0002659-2, Sondicator Probes". United States. doi:10.2172/888621. https://www.osti.gov/servlets/purl/888621.
@article{osti_888621,
title = {Performance, Thermal, and Vibration Qualification Testing of Zetec Acoustic Transducers, Model Z0002659-2, Sondicator Probes},
author = {Jacobson, G and Gemberling, S and Lavietes, A},
abstractNote = {This report is a result of Qualification Test Plan No.001 prepared by Anthony Lavietes. The Qualification Test Plan outlines a list of requirements for thermal and vibrational testing of Zetac's Z0002659-2 Sondicator Probe acoustic transducers (hereafter called ''transducers''). The Zetec transducers are used in a system that employs an array of 7 acoustic transducers. Qualification testing of these transducers was required since they are a modified version of a standard catalog item from the manufacturer. This report documents the thermal, vibrational, and performance testing that was performed on a sampling of these transducers in order to qualify them for flight. A total of 14 transducers were tested. All 14 passed qualification testing with no failures.},
doi = {10.2172/888621},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Fri Mar 10 00:00:00 EST 2006},
month = {Fri Mar 10 00:00:00 EST 2006}
}

Technical Report:

Save / Share: