Diagnostics for the LCLS Photoinjector Beamline
Two spectrometers have been added to the LCLS photoinjector beamline. The first one will be located close to the exit of the Photoinjector RF gun. With this diagnostic, we will measure beam energy, energy spread (correlated and uncorrelated), possibly deleterious structure in the longitudinal phase space induced by longitudinal space charge force, and slice thermal emittance ... This extensive characterization of the 5MeV electron bunch will be made possible by combining this spectrometer with other diagnostics (YAG screens and Cerenkov Radiator). A second spectrometer located at the end of the beamline has been designed to characterize the 6 dimensional phase space of the 135MeV beam to be injected in the main accelerator. At that second spectrometer station, we will measure energy, energy spread (correlated and uncorrelated), longitudinal phase space, slice emittances ... Those last two measurements require using this spectrometer in combination with the transverse RF deflecting cavity and with the quadrupole scan emittance station. The designs of these two spectrometers have been supported by simulations from MAD and PARMELA.