Comparison of 3D Maximum A Posteriori and Filtered Backprojection algorithms for high resolution animal imaging in microPET

Journal Article ·
OSTI ID:843118
We have evaluated the performance of two three dimensional reconstruction algorithms with data acquired from microPET, a high resolution tomograph dedicated to small animal imaging. The first was a linear filtered-backprojection algorithm (FBP) with reprojection of the missing data and the second was a statistical maximum-aposteriori probability algorithm (MAP). The two algorithms were evaluated in terms of their resolution performance, both in phantoms and in vivo. Sixty independent realizations of a phantom simulating the brain of a baby monkey were acquired, each containing 3 million counts. Each of these realizations was reconstructed independently with both algorithms. The ensemble of the sixty reconstructed realizations was used to estimate the standard deviation as a measure of the noise for each reconstruction algorithm. More detail was recovered in the MAP reconstruction without an increase in noise relative to FBP. Studies in a simple cylindrical compartment phantom demonstrated improved recovery of known activity ratios with MAP. Finally in vivo studies also demonstrated a clear improvement in spatial resolution using the MAP algorithm. The quantitative accuracy of the MAP reconstruction was also evaluated by comparison with autoradiography and direct well counting of tissue samples and was shown to be superior.
Research Organization:
Ernest Orlando Lawrence Berkeley National Laboratory, Berkeley, CA (US)
Sponsoring Organization:
USDOE Director. Office of Science; National Cancer Institute Grant R01 CA69370 and R01 CA59794
DOE Contract Number:
AC03-76SF00098
OSTI ID:
843118
Report Number(s):
LBNL--45516
Country of Publication:
United States
Language:
English

Similar Records

Filtered backprojection proton CT reconstruction along most likely paths
Journal Article · Fri Mar 15 00:00:00 EDT 2013 · Medical Physics · OSTI ID:22130540

Noise properties of filtered-backprojection and ML-EM reconstructed emission tomographic images
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6057481

Evaluation of the spline reconstruction technique for PET
Journal Article · Tue Apr 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22250870