Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Opticla and X-Ray Streak Camera Gain Measurements

Journal Article · · Review of Scientiifc Instruments
DOI:https://doi.org/10.1063/1.1787929· OSTI ID:833598
 [1]; ;
  1. Laboratory for Laser Energetics, University of Rochester, Rochester, NY

Measurements of streak camera gain as the number of CCD (charge-coupled-device) electrons recorded per single-electron events hitting the streak tube phosphor are presented. The CCD is fiber optically coupled to the streak tube output; there is no image intensifier in the system. The gain is measured from the signal-to-noise (SNR) ratio of the recorded photoelectrons. This technique allows us to verify that the photoelectron SNR follows Poisson statistics and to establish the linear dynamic range.

Research Organization:
Laboratory for Laser Energetics
Sponsoring Organization:
USDOE
DOE Contract Number:
FC52-92SF19460
OSTI ID:
833598
Report Number(s):
DOE/SF-19460-561; 1484; 2003-196
Journal Information:
Review of Scientiifc Instruments, Journal Name: Review of Scientiifc Instruments Journal Issue: 10 Vol. 75; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Streak camera dynamic range optimization
Conference · Tue Sep 01 00:00:00 EDT 1987 · OSTI ID:5887465

Streak Camera Performance with Large-Format CCD Readout
Conference · Tue Jul 08 00:00:00 EDT 2003 · OSTI ID:15005836

Comparison of Streak Tube Performance
Technical Report · Wed Nov 03 23:00:00 EST 2004 · OSTI ID:15011417