New approach to jitter reduction of an x-ray streak camera in accumulation mode.

Conference ·
DOI:https://doi.org/10.1117/12.463384· OSTI ID:797920

An x-ray streak camera operating in accumulation mode was developed for studying ultrafast dynamics at synchrotron facilities. A laser-triggered photoconductive switch was used as a sweeping unit to obtain low timing jitter. The fast rise time of the ramp pulse generated by the switch (90 ps) combined with the fast response of the traveling wave deflection plates (150 ps) significantly reduced the jitter caused by the shot-to-shot laser fluctuation. At {approx}1% rms (root mean square) laser energy fluctuation, the resolution of the camera is 1.1 ps when over 5000 laser shots were accumulated. This is two times better than that of the previous design with slower response (300 ps) deflection plates.

Research Organization:
Argonne National Lab., IL (US)
Sponsoring Organization:
US Department of Energy (US)
DOE Contract Number:
W-31109-ENG-38
OSTI ID:
797920
Report Number(s):
ANL/XFD/CP-108269
Country of Publication:
United States
Language:
English

Similar Records

An accumulative x-ray streak camera with sub-600-fs temporal resolution and 50-fs timing jitter
Journal Article · Tue Dec 31 23:00:00 EST 2002 · Applied Physics Letters · OSTI ID:15003796

Signal averaging x-ray streak camera with picosecond jitter
Journal Article · Thu Feb 29 19:00:00 EST 1996 · Review of Scientific Instruments · OSTI ID:279263

Synchroscan streak camera imaging at a 15-MeV photoinjector with emittance exchange
Journal Article · Wed Feb 29 23:00:00 EST 2012 · Submitted to Nucl.Instrum.Meth.A. · OSTI ID:1038936