Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Application of a Bender-Focuser for Cold Neutron PGAA

Conference ·
OSTI ID:786614
A bender-focuser has been installed and characterized at the Cold Neutron Prompt Gamma Activation Analysis facility. It accepts the neutron beam exiting the {sup 58}Ni-coated guide, 45 x 50 mm in cross section, and compresses it to a sub-millimetre focus 42.5 mm below the axis of the direct beam. The gain in neutron intensity at the 0.65-mm FWHM spot is 20. However, by placing the lens in the beam, the overall gamma background has increased. Compared with a previously designed lens with the focus on the beam axis, the bender-focuser results in less increase in the gamma background. We report a work in progress using the bender-focuser for material studies.
Research Organization:
National Institute of Standards and Technology, Gaithersburg, MD (US)
Sponsoring Organization:
none (US)
OSTI ID:
786614
Report Number(s):
none; ISSN 0003-018X; CODEN TANSAO; ISSN 0003-018X; CODEN TANSAO
Country of Publication:
United States
Language:
English