X-ray BPM-based feedback system at the APS storage ring
At the Advanced Photon Source (APS) storage ring, the X-ray beam position monitors (X-BPMs) measure accurate photon position down to the submicron level. This level of stable measurement has been possible due to (1) superior thermal insulation and vibration damping of the X-ray BPM support structure, (2) minimal dependence on the bunch pattern and intensity variations, and (3) use of ultrastable preamplifiers and processing electronics. A new X-BPM interface is under development and will be discussed here. This interface will be integrated into the existing rf-based orbit feedback systems. To study preliminary results, an experimental X-BPM orbit feedback set-up was developed and implemented in one of the bending magnet beamlines. The results from this set-up are encouraging. For an operational fill, a typical orbit drift of 30 microns (at X-ray BPMs) has been reduced to less than 5 microns. The fill-to-fill photon orbit reproducibility has been improved from 75 microns to less than 10 microns.