Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

SQUID instruments and shielding for low-level magnetic measurements

Journal Article · · J. Appl. Phys.; (United States)
DOI:https://doi.org/10.1063/1.323659· OSTI ID:7324499

We have used a thick-walled aluminum enclosure to provide low-frequency shielding for biomagnetic studies with SQUID magnetic gradiometers. The shield provides an attenuation proportional to frequency above 1/3 Hz. A 24-hole fractional-turn SQUID gradiometer has been used in the enclosure to carry out studies of magnetoencephalography on humans. An ''asymmetric'' configuration for a second-derivative gradiometer has been developed which provides a significant increase in sensitivity and resolution over more conventional symmetric arrays. (AIP)

Research Organization:
Cryogenics Division, Institute for Basic Standards, National Bureau of Standards, Boulder, Colorado 80302
OSTI ID:
7324499
Journal Information:
J. Appl. Phys.; (United States), Journal Name: J. Appl. Phys.; (United States) Vol. 48:2; ISSN JAPIA
Country of Publication:
United States
Language:
English

Similar Records

Base distance optimization for SQUID gradiometers
Conference · Wed Dec 30 23:00:00 EST 1998 · OSTI ID:350912

Multichannel DC SQUID sensor array for biomagnetic applications
Conference · Thu Feb 28 23:00:00 EST 1991 · IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6088665

Thermal magnetic noise generated by an eddy current shielded room
Journal Article · Mon Mar 31 23:00:00 EST 1986 · Rev. Sci. Instrum.; (United States) · OSTI ID:6113993