Evaluation of factors contributing to position accuracy in delay line readout of MWPC

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:7307389

A method to accurately measure the resolution of multiwire proportional chambers with delay line readout is described. The contribution of signal to noise, pulse shape, pulse rise time and discriminator slewing to the position accuracy is presented. Results show that position resolution better than 100 microns is possible.

Research Organization:
Univ. of California, Berkeley
OSTI ID:
7307389
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Journal Name: IEEE Trans. Nucl. Sci.; (United States) Vol. NS-24:1; ISSN IETNA