skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Comparative analysis of dioxins and furans in ambient air by high-resolution and electron-capture mass spectrometry

Technical Report ·
OSTI ID:7264138

Known mixtures and unknown atmospheric sample extracts containing polychlorinated dibenzo-p-dioxins and polychlorinated dibenzofurans (PCDD/F) were analyzed by both electron impact, high resolution, mass spectrometry (HRMS) and by electron capture, negative ion, low resolution mass spectrometry (ECNI). PCDD/F concentrations measured by the two methods were comparable, typically agreeing with + or - 33%. The major difference between the two techniques is that HRMS easily detects 2,3,7,8- tetrachlorodibenzo-p-dioxin but ECNI does not. Results suggest that ECNI can be a sensitive low cost alternative to HRMS for the determination of PCDD/F concentrations.

Research Organization:
Indiana Univ., Bloomington, IN (United States)
OSTI ID:
7264138
Report Number(s):
PB-92-206440/XAB; CNN: EPA-R-812588
Resource Relation:
Other Information: Pub. in Chemosphere, v24 n4 p421-426 1992. Sponsored by Environmental Protection Agency, Research Triangle Park, NC. Atmospheric Research and Exposure Assessment Lab
Country of Publication:
United States
Language:
English