Title: Mercuric iodide photodetector-cesium iodide scintillator gamma ray spectrometers

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:7254475

Scintillation spectrometers with mercuric iodide as the photodetector were evaluated for their spectral response to gammas in the energy range of 60 keV to 1332 keV. A /sup 137/Cs (662 keV) photopeak resolution of better than 5% FWHM was obtained due to the low-noise characteristics and high-quantum efficiency of the mercuric iodide photodetector. Photopeak broadening due to the scintillator, electronic noise, and statistical noise versus gamma energy are detailed.

Research Organization:
EG and G Energy Measurements, Inc., Santa Barbara Operations, Goleta, CA (US)
OSTI ID:
7254475
Report Number(s):
CONF-871006-
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Vol. 35:1; Conference: 34. nuclear science symposium and 19. nuclear power systems symposium, San Francisco, CA, USA, 21 Oct 1987
Country of Publication:
United States
Language:
English