Procedure manual for the estimation of average indoor radon-daughter concentrations using the Electronic Radon Progeny Integrating Sampling Unit (ERPISU) method
Technical Report
·
OSTI ID:7242948
One of the remedial action measurements needs is the estimation of average indoor radon-daughter concentration (RDC). One method for accomplishing such estimations is the Electric Radon Progeny Integrating Sampling Unit (ERPISU) method. Electronic RPISUs differ from standard RPISUs in that an Electronic RPISU measures radon-daughter concentrations using a solid-state alpha detector, rather than the thermoluminescent dosimeter employed by a standard RPISU. This procedure manual describes hoe to set up and operate a system for obtaining ERPISU measurements, and how to derive annual average RDC estimates from those measurements as well as RDC estimates for comparsion against selected standards. This manual contains both procedures specific to the Elberine WLM-1A Working Level Monitor and general guidelines applicable to any type of Electronic RPISU that may be approved by DOE. Necessary equipment items include samplers, readout units or computer interfaces, filters, radioactive alpha standard, and an airflow measuring device. The annula average radon-daughter concentration is estimated as the mean of six such measurements, each approximately one week in duration and spaced approximately two months apart. Criteria for making fewer than six measurements, including the abbreviated RPISU method, and for handling missed or invalid measurements are also presented. 9 refs., 4 figs., 2 tabs.
- Research Organization:
- UNC Geotech, Grand Junction, CO (USA)
- DOE Contract Number:
- AC07-86ID12584
- OSTI ID:
- 7242948
- Report Number(s):
- DOE/ID/12584-15; UNC/GJ-39(TMC); ON: DE88008583
- Country of Publication:
- United States
- Language:
- English
Similar Records
Procedure manual for the estimation of average indoor radon-daughter concentrations using the Radon Progeny Integrating Sampling Unit (RPISU) method
Field study of indoor average radon-daughter estimation methods
Procedure manual for the estimation of average indoor radon-daughter concentrations using the filtered alpha-track method
Technical Report
·
Fri Jan 31 23:00:00 EST 1986
·
OSTI ID:6081937
Field study of indoor average radon-daughter estimation methods
Technical Report
·
Fri Aug 01 00:00:00 EDT 1986
·
OSTI ID:5054104
Procedure manual for the estimation of average indoor radon-daughter concentrations using the filtered alpha-track method
Technical Report
·
Thu Mar 31 23:00:00 EST 1988
·
OSTI ID:5110968
Related Subjects
500300* -- Environment
Atmospheric-- Radioactive Materials Monitoring & Transport-- (-1989)
54 ENVIRONMENTAL SCIENCES
AIR SAMPLERS
ALPHA DECAY RADIOISOTOPES
ALPHA DETECTION
CALIBRATION
CHARGED PARTICLE DETECTION
DAUGHTER PRODUCTS
DAYS LIVING RADIOISOTOPES
DETECTION
ECOLOGICAL CONCENTRATION
EQUIPMENT
EVEN-EVEN NUCLEI
HEAVY NUCLEI
ISOTOPES
MILL TAILINGS
NUCLEI
PORTABLE EQUIPMENT
RADIATION DETECTION
RADIOECOLOGICAL CONCENTRATION
RADIOISOTOPES
RADON 222
RADON ISOTOPES
REMEDIAL ACTION
SAMPLERS
SOLID WASTES
TAILINGS
WASTES
Atmospheric-- Radioactive Materials Monitoring & Transport-- (-1989)
54 ENVIRONMENTAL SCIENCES
AIR SAMPLERS
ALPHA DECAY RADIOISOTOPES
ALPHA DETECTION
CALIBRATION
CHARGED PARTICLE DETECTION
DAUGHTER PRODUCTS
DAYS LIVING RADIOISOTOPES
DETECTION
ECOLOGICAL CONCENTRATION
EQUIPMENT
EVEN-EVEN NUCLEI
HEAVY NUCLEI
ISOTOPES
MILL TAILINGS
NUCLEI
PORTABLE EQUIPMENT
RADIATION DETECTION
RADIOECOLOGICAL CONCENTRATION
RADIOISOTOPES
RADON 222
RADON ISOTOPES
REMEDIAL ACTION
SAMPLERS
SOLID WASTES
TAILINGS
WASTES