Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Evaluation of long life (L/sup 2/) microchannel plates for x-ray photon counting

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:7198158
The authors present an evaluation of Galileo Long-Life (L/sup 2/) microchannel plates (MCPs) operated in pulse counting mode. Modal gain, pulse height FWHM, gain uniformity and high count rate performance data are compared with measurements made with MCPs of conventional glass composition. Despite the absence of potassium (and hence, of the ..beta..-emitter /sup 40/K) from the L/sup 2/ glass, dark count rates for the L/sup 2/ MCPs are rather higher (times --5) than those of standard MCPs, due to the presence in the new glass of the radioisotope /sup 87/Rb. The manufacturer's key claim (made for the linear mode of plate operation), that the gain of L/sup 2/ MCPs falls much less rapidly than that of standard plates, is examined in the light of the results from a number of gain fatigue lifetests.
Research Organization:
X-ray Astronomy Group, Dept. of Physics, Univ. of Leicester, Leicester LE1 7RH (GB)
OSTI ID:
7198158
Report Number(s):
CONF-871006-
Conference Information:
Journal Name: IEEE Trans. Nucl. Sci.; (United States) Journal Volume: 35:1
Country of Publication:
United States
Language:
English