Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Pulse filtering for thick mercuric iodide detectors

Conference ·
OSTI ID:7194140
We have investigated a number of electronic pulse filtering techniques for thick mercuric iodide detectors and have achieved improved spectral response for detectors of 1 to 5 mm thickness and gamma ray energies up to 1.5 MeV. Conventional gaussian and gated integration filters were considered as well as more sophisticated filter combinations in order to develop methods of reducing charge measurement errors caused by poor hole collection. When using these techniques, energy resolutions of 3 to 6% and peak-to-valley ratios of 5:1 are common. In this paper we discuss these approaches and present results with thick detectors.
Research Organization:
EG and G Energy Measurements, Inc., Goleta, CA (USA). Santa Barbara Operations
DOE Contract Number:
AC08-83NV10282
OSTI ID:
7194140
Report Number(s):
EGG-10282-2097; CONF-861007-2; ON: DE87001782
Country of Publication:
United States
Language:
English