Gas chromatographic method for measuring nitrogen dioxide and peroxyacetyl nitrate in air without compressed gas cylinders
A gas chromatographic technique that measures atmospheric concentrations of peroxyacetyl nitrate (PAN) and NO/sub 2/ has been developed that uses luminol-based chemiluminescence for detection. The carrier gas is air that has been scrubbed by passing it over FeSO/sub 4/, which eliminates the need for any compressed gas cylinders. A novel gas sampling system and time enable variable sample volumes of contaminated air to be injected. Ambient PAN and NO/sub 2/ measurements can be made every 40 s with detection limits of 0.12 ppb for PAN and 0.2 ppb for NO/sub 2/. Seven other atmospheric species, including ozone, gave no interference. Linear response was observed for NO/sub 2/ from 0.2 to 170 ppb and for PAN from 1 to 70 ppb.
- Research Organization:
- Univ. of Denver, CO (USA)
- OSTI ID:
- 7144097
- Journal Information:
- Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 60:8; ISSN ANCHA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Capillary gas chromatographic analysis of pans with luminol chemilumnescent detection
Capillary gas chromatographic analysis of nitrogen dioxide and pans with luminol chemiluminescent detection.
Related Subjects
400102* -- Chemical & Spectral Procedures
500200 -- Environment
Atmospheric-- Chemicals Monitoring & Transport-- (-1989)
54 ENVIRONMENTAL SCIENCES
AIR
AIR SAMPLERS
AMINES
AZINES
CHALCOGENIDES
CHEMICAL ANALYSIS
CHEMILUMINESCENCE
CHROMATOGRAPHY
DATA
EQUIPMENT
ESTERS
EXPERIMENTAL DATA
FLUIDS
GAS CHROMATOGRAPHY
GASES
HETEROCYCLIC COMPOUNDS
INFORMATION
LUMINESCENCE
LUMINOL
NITRATES
NITRIC ACID ESTERS
NITROGEN COMPOUNDS
NITROGEN DIOXIDE
NITROGEN OXIDES
NUMERICAL DATA
ORGANIC COMPOUNDS
ORGANIC NITROGEN COMPOUNDS
OXIDES
OXYGEN COMPOUNDS
PEROXYACETYL NITRATE
PHTHALAZINES
PYRIDAZINES
QUANTITATIVE CHEMICAL ANALYSIS
QUANTITY RATIO
SAMPLERS
SEPARATION PROCESSES
TRACE AMOUNTS