Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Mailed TL dosimeters for monitoring the output from diagnostic x-ray equipment

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.595457· OSTI ID:7136576
Mailing thermoluminescent (TL) dosimeters for exposure to users of diagnostic x-ray equipment is a simple and inexpensive way of monitoring output from x-ray units. The readings given by TL dosimeters were originally compared with ionization chamber measurements and shown to give reliable and compatible results. Mailed TL dosimeters alone were later used in a university hospital to check the functioning of exposure timers in thorax and Bucky stands. A thorough survey of the outputs from all x-ray units in use throughout the country can be performed within a few months using this technique.
Research Organization:
Institute of Radiation Protection, Helsinki, Finland
OSTI ID:
7136576
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 11:1; ISSN MPHYA
Country of Publication:
United States
Language:
English

Similar Records

Mailed thermoluminescent dosimeter determination of entrance skin exposure and half-value layer in mammography
Journal Article · Sun Oct 01 00:00:00 EDT 1978 · AJR, Am. J. Roentgenol.; (United States) · OSTI ID:5502482

Evaluation of a commercial diode monitor for mailed quality control of therapy units
Journal Article · Mon Oct 01 00:00:00 EDT 1990 · International Journal of Radiation Oncology, Biology and Physics; (USA) · OSTI ID:6141349

Linear accelerator calibration monitor with a memory
Journal Article · Sat Sep 01 00:00:00 EDT 1979 · Med. Phys.; (United States) · OSTI ID:5893223