Uniform magnetic fields and double-wrapped coil systems: Improved techniques for the design of bioelectromagnetic experiments
- California Inst. of Technology, Pasadena (United States)
A common mistake in biomagnetic experimentation is the assumption that Helmholtz coils provide uniform magnetic fields; this is true only for a limited volume at their center. Substantial improvements on this design have been made during the past 140 years with systems of three, four, and five coils. Numerical comparisons of the field uniformity generated by these designs are made here, along with a table of construction details and recommendations for their use in experiments in which large volumes of uniform intensity magnetic exposures are needed. Double-wrapping, or systems of bifilar windings, can also help control for the non-magnetic effects of the electric coils used in many experiments. In this design, each coil is wrapped in parallel with two separate, adjacent strands of copper wire, rather than the single strand used normally. If currents are flowing in antiparallel directions, the magnetic fields generated by each strand will cancel and yield virtually no external magnetic field, whereas parallel currents will yield an external field. Both cases will produce similar non-magnetic effects of ohmic heating, and simple measures can reduce the small vibration and electric field differences. Control experiments can then be designed such that the only major difference between treated and untreated groups is the presence or absence of the magnetic field. Double-wrapped coils also facilitate the use of truly double-blind protocol, as the same apparatus can be used either for experimental or control groups.
- OSTI ID:
- 7120086
- Journal Information:
- Bioelectromagnetics (New York); (United States), Vol. 13:5; ISSN 0197-8462
- Country of Publication:
- United States
- Language:
- English
Similar Records
Conceptual design of a hollow cable conductor for the Large Coil Program
Generation of Uniform Magnetic Fields by Means of Air-Core Coils
Related Subjects
42 ENGINEERING
ELECTRIC COILS
ELECTROMAGNETIC FIELDS
LABORATORY EQUIPMENT
DESIGN
ENVIRONMENTAL EXPOSURE
MAGNETIC FIELDS
RADIOBIOLOGY
BIOLOGY
ELECTRICAL EQUIPMENT
EQUIPMENT
560400* - Other Environmental Pollutant Effects
420200 - Engineering- Facilities
Equipment
& Techniques