Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

The Fixed-Cavity VISAR

Technical Report ·
OSTI ID:7115492
A designed and assembled method for a non-adjustable Interferometer cavity has been developed at Sandia National Laboratories which has enabled the development of a Fixed-Cavity Velocity Interferometer System for Any Reflector (VISAR). In this system, the critical interference adjustments are performed during assembly of the interferometer cavity, freeing the user from an otherwise repetitive task. The Fixed-Cavity VISAR System is constructed in modular form. Compared to previous VISAR systems, it is easy to use, and gives high quality results. 6 refs.
Research Organization:
Sandia National Labs., Albuquerque, NM (United States)
Sponsoring Organization:
DOE; USDOE, Washington, DC (United States)
DOE Contract Number:
AC04-76DP00789
OSTI ID:
7115492
Report Number(s):
SAND-92-0162; ON: DE92019325
Country of Publication:
United States
Language:
English