Method of test for radioactive tritium in water - 1973

Book ·
OSTI ID:7101228

A method is described for the liquid scintillation measurement of tritium radioactivity in water, T/sub 2/O or HTO. The minimum detectable activity is 2pCi/ml for a 50-ml sample counted for 20 min, in a counter where the background is 30 cpm and the efficiency is 13.5%. Multiple solutes are used in the scintillator to provide the best combination of wavelength and pulse height for this application.