Radiation exposure to angiographer during coronary arteriography using the U arm image amplifier
Journal Article
·
· Cathet. Cardiovasc. Diagn.; (United States)
Radiation exposure to the angiographer was measured using thermoluminescent dosimeters (TLD), commercial film badges, and appropriate ionization survey meters. Data were collected during 47 coronary artery examinations with the lead overleaf. During a single coronary examination, the angiographer receives 0.9 mR to the lens and 2.8 mR to the hand. The calculated yearly exposure to the angiographer performing 10 examinations per week is 450 mR to the lens, 9% of the maximum permissible dose (MPD), and 1400 mR to the hand, 2% of the MPD. Exposure rates for the left anterior oblique (LAO) and right anterior oblique (RAO) projections during cine and fluoroscopy were taken at the position occupied by the angiographer. Scatter radiation is 5-10 times greater in the LAO projection. The second trial was monitored during 18 coronary artery examinations minus the lead overleaf. A significant decrease in radiation exposure was accomplished through the use of the lead overleaf.
- Research Organization:
- Allegheny General Hospital, Pittsburgh
- OSTI ID:
- 7086752
- Journal Information:
- Cathet. Cardiovasc. Diagn.; (United States), Journal Name: Cathet. Cardiovasc. Diagn.; (United States) Vol. 1:3; ISSN CCDID
- Country of Publication:
- United States
- Language:
- English
Similar Records
Gated cardiac tomographic visualization of coronary artery calcification
Coronary arteriography and left ventriculography during spontaneous and exercise-induced ST segment elevation in patients with variant angina
Anomalous origin of the left coronary artery: angiographic and myocardial perfusion scintigraphic correlates
Journal Article
·
Sun Jul 01 00:00:00 EDT 1984
· Radiology; (United States)
·
OSTI ID:6659942
Coronary arteriography and left ventriculography during spontaneous and exercise-induced ST segment elevation in patients with variant angina
Journal Article
·
Thu Sep 01 00:00:00 EDT 1983
· Am. Heart J.; (United States)
·
OSTI ID:5477178
Anomalous origin of the left coronary artery: angiographic and myocardial perfusion scintigraphic correlates
Journal Article
·
Tue Mar 31 23:00:00 EST 1981
· Chest; (United States)
·
OSTI ID:5504496
Related Subjects
560151* -- Radiation Effects on Animals-- Man
63 RADIATION, THERMAL, AND OTHER ENVIRON. POLLUTANT EFFECTS ON LIVING ORGS. AND BIOL. MAT.
BIOMEDICAL RADIOGRAPHY
DIAGNOSTIC TECHNIQUES
DOSES
MEDICINE
MONITORING
NUCLEAR MEDICINE
PERSONNEL
RADIATION DOSES
RADIATION MONITORING
RADIATION PROTECTION
RADIOLOGICAL PERSONNEL
RADIOLOGY
63 RADIATION, THERMAL, AND OTHER ENVIRON. POLLUTANT EFFECTS ON LIVING ORGS. AND BIOL. MAT.
BIOMEDICAL RADIOGRAPHY
DIAGNOSTIC TECHNIQUES
DOSES
MEDICINE
MONITORING
NUCLEAR MEDICINE
PERSONNEL
RADIATION DOSES
RADIATION MONITORING
RADIATION PROTECTION
RADIOLOGICAL PERSONNEL
RADIOLOGY