Monte Carlo evaluation of specific dose constants in water for /sup 125/I seeds

Journal Article · · Med. Phys.; (United States)
OSTI ID:7063843
To clinically utilize relative three-dimensional dose distribution data for /sup 125/I seeds in water, the specific dose constant must be known. This constant is the ratio of absolute dose rate in medium, 1 cm from the source on its transverse bisector, to source strength. We have used the Monte Carlo method to calculate specific dose constants in water for three types of /sup 125/I seeds (models 6711, 6701, and 6702). Our photon transport code realistically models the complex internal structure of these interstitial sources as well as the exposure standardization experiments conducted at the National Bureau of Standards which form the basis of /sup 125/I source strength specification. In addition, relative attenuation factors for the three seed types are presented. Our results indicate that published data overestimate the specific dose constant by 10% to 14%.
Research Organization:
Department of Radiation Oncology and the Arizona Cancer Center, University of Arizona, Tucson, Arizona 85724
OSTI ID:
7063843
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 15:5; ISSN MPHYA
Country of Publication:
United States
Language: