An x-ray technique for precision laser beam synchronization
Journal Article
·
· Review of Scientific Instruments; (United States)
- Lawrence Livermore National Laboratory, P. O. Box 5508, Livermore, California 94550 (United States)
A new x-ray technique for recording the relative arrival times of multiple laser beams at a common target with better than [plus minus]10-ps accuracy has been implemented at the Nova laser facility. 80-ps, 3[omega] Nova beams are focused to separate locations on a gold ribbon target viewed from the side. The measurement consists of using well-characterized reentrant x-ray streak cameras for one-dimensional streaked imaging of the [gt]3-keV x rays emanating from these isolated laser plasmas. After making the necessary corrections for the differential laser, x-ray and electron transit times involved, timing offsets as low as [plus minus]7 ps are resolved, and on subsequent shots, corrected for, verified, and independently checked. This level of synchronization proved critical in meeting the power balance requirements for indirectly driven pulse-shaped Nova implosions.
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 7060335
- Journal Information:
- Review of Scientific Instruments; (United States), Journal Name: Review of Scientific Instruments; (United States) Vol. 66:1; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
An x-ray technique for precision laser beam synchronization
Timing between streak cameras with a precision of 10 ps
Precision Nova operations
Conference
·
Mon May 02 00:00:00 EDT 1994
·
OSTI ID:10173008
Timing between streak cameras with a precision of 10 ps
Conference
·
Thu Dec 06 23:00:00 EST 1990
·
OSTI ID:6428341
Precision Nova operations
Conference
·
Fri Sep 01 00:00:00 EDT 1995
·
OSTI ID:109688
Related Subjects
70 PLASMA PHYSICS AND FUSION TECHNOLOGY
700320* -- Plasma Diagnostic Techniques & Instrumentation-- (1992-)
CAMERAS
DESIGN
DETECTION
ENERGY RANGE
KEV RANGE
KEV RANGE 01-10
MEASURING INSTRUMENTS
NOVA FACILITY
PERFORMANCE
PLASMA DIAGNOSTICS
RADIATION DETECTION
RADIATION DETECTORS
RESOLUTION
STREAK CAMERAS
TIME RESOLUTION
TIMING PROPERTIES
X-RAY DETECTION
700320* -- Plasma Diagnostic Techniques & Instrumentation-- (1992-)
CAMERAS
DESIGN
DETECTION
ENERGY RANGE
KEV RANGE
KEV RANGE 01-10
MEASURING INSTRUMENTS
NOVA FACILITY
PERFORMANCE
PLASMA DIAGNOSTICS
RADIATION DETECTION
RADIATION DETECTORS
RESOLUTION
STREAK CAMERAS
TIME RESOLUTION
TIMING PROPERTIES
X-RAY DETECTION