Spectroscopic diagnostics of tritium recycling in TFTR

Journal Article · · Review of Scientific Instruments; (United States)
DOI:https://doi.org/10.1063/1.1146315· OSTI ID:7060021
; ; ;  [1]
  1. Princeton University, Plasma Physics Laboratory, P. O. Box 451, Princeton, New Jersey 08543 (United States)
Spectroscopic measurements of tritium Balmer-alpha (T[sub [alpha]]) emission from a fusion plasma are presented. A Fabry--Perot interferometer is used to measure the H[sub [alpha]], D[sub [alpha]], T[sub [alpha]] spectrum in the current D--T experimental campaign on TFTR and the contributions of H, D, and T are separated by spectral analysis. The T[sub [alpha]] line was measurable at concentrations T[sub [alpha]] /(H[sub [alpha]]+D[sub [alpha]]+T[sub [alpha]]) down to 2%.
DOE Contract Number:
AC02-76CH03073
OSTI ID:
7060021
Journal Information:
Review of Scientific Instruments; (United States), Journal Name: Review of Scientific Instruments; (United States) Vol. 66:1; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States