Helium-to-neon modification for a DuPont 120SSA or 24-120 leak detector
When helium leak detecting is hindered by a system or atmosphere saturated with helium, leak checking can be continued by modifying the DuPont 120SSA/24-120 Mass Spectrometer Leak Detector to detect neon. Although DuPont markets a three gas leak detector (helium, neon and argon), the DuPont 120SSA/24-120 is convertible to neon by simply changing the magnet and retuning for neon peak. The sensitivity is 2 x 10/sup -9/ Atm/cc/sec/division neon. If an increase in sensitivity is desired, throttle diffusion pump with valve or install choke. Modification procedures and tuning procedures are described. (WHK)
- Research Organization:
- Fermi National Accelerator Lab., Batavia, IL (USA)
- DOE Contract Number:
- AC02-76CH03000
- OSTI ID:
- 7057378
- Report Number(s):
- FERMILAB/TM-1072; ON: DE83000617