A backprojection filtering algorithm for a spatially varying focal length collimator

Journal Article · · IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States)
DOI:https://doi.org/10.1109/42.310886· OSTI ID:6977488
;  [1]
  1. Univ. of Utah, Salt Lake City, UT (United States). Dept. of Radiology

Fan-beam collimators are used in single photon emission computed tomography to improve the sensitivity for imaging of small organs. The disadvantage of fan-beam collimation is the truncation of projection data surrounding the organ of interest or, in those cases of imaging large patients, of the organ itself producing reconstruction artifacts. A spatially varying focal length fan-beam collimator has been proposed to eliminate the truncation problem and to maintain good sensitivity for the organ of interest. The collimator is constructed so that the focal lengths of the holes vary across the face of the collimator with the shortest focal lengths at the center and the longer focal lengths at the periphery of the collimator. The variation of the focal length can have various functional forms but in the authors' work it is assumed to increase monotonically toward the edge of the collimator. A backprojection filtering reconstruction algorithm is derived for this type of collimation. The algorithm first backproject the projections, then perform a two-dimensional filtering. The algorithm is efficient and has been tested via computer simulations.

OSTI ID:
6977488
Journal Information:
IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States), Journal Name: IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States) Vol. 13:3; ISSN 0278-0062; ISSN ITMID4
Country of Publication:
United States
Language:
English

Similar Records

Fan-beam reconstruction algorithm for a spatially varying focal length collimator
Journal Article · Wed Sep 01 00:00:00 EDT 1993 · IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5641317

Backprojection filtering for variable orbit fan-beam tomography
Journal Article · Tue Aug 01 00:00:00 EDT 1995 · IEEE Transactions on Nuclear Science · OSTI ID:148077

A multifocal collimator with circularly distributed focal points for SPECT imaging
Conference · Mon Aug 01 00:00:00 EDT 1994 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6838684