Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Membrane electrode for the determination of actinyl(VI) cations

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00253a040· OSTI ID:6972601
A coated wire specific electrode has been developed for actinyl(VI) cations. The electrode responds in Nernstian fashion to actinyl(VI) concentrations of 10/sup -5/ - 10/sup -2/ M between pH 2 and 5. The interference by M(I), M(II), M(III), and MO/sub 2//sup +/ cations is small even when (M(III)):)An(anion)O/sub 2//sup 2 +/) approx. = 10. However, Th(IV) does interfere. The electrode can be used in complexation and redox studies of UO/sub 2//sup 2 +/, NpO/sub 2//sup 2 +/, and PuO/sub 2//sup 2 +/ within its useful range of pH and AnO/sub 2//sup 2 +/ concentration.
Research Organization:
Florida State Univ., Tallahassee
OSTI ID:
6972601
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 55:2; ISSN ANCHA
Country of Publication:
United States
Language:
English

Similar Records

Unexpected Actinyl Cation-Directed Structural Variation in Neptunyl(VI) A-Type Tri-lacunary Heteropolyoxotungstate Complexes
Journal Article · Tue Apr 21 20:00:00 EDT 2015 · Inorganic Chemistry · OSTI ID:1239107

Actinide cation-cation complexes
Thesis/Dissertation · Wed Nov 30 23:00:00 EST 1994 · OSTI ID:34204

Actinyl cation–cation interactions in the gas phase: an accurate thermochemical study