Analytical laboratories method No. 4002 - the radiometric determination of radium-226 and radium-228 in various matrices

Technical Report ·
OSTI ID:6951700
This method is applicable to the analysis of sump effluents and production by-product materials that are soluble in mineral acids. This method is used to separate, and analyze, Ra-226 and Ra-228 from solutions containing uranium, thorium, decay products, and nonradioactive materials. Typical analysis time for both Ra-226 and Ra-228 is three days. The radium isotopes can be determined to a lower detection limits of 0.010 d/m-mL.
Research Organization:
Westinghouse Materials Co., Fernald, OH (USA)
DOE Contract Number:
AC05-86OR21600
OSTI ID:
6951700
Report Number(s):
DOE/OR/21600-T20; ON: DE88012268
Country of Publication:
United States
Language:
English