Intense ion beam diagnostics for particle beam fusion experiments on PBFA II (invited)

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1140089· OSTI ID:6949782
A review of the diagnostics used at Sandia National Laboratories to measure the parameters of intense proton and lithium beams generated on the PBFA-II accelerator will be presented. These diagnostics consist of several types, namely: K..cap alpha.. x-ray pinhole cameras, a multiframe dE/dx ion pinhole camera, a p-i-n array ion pinhole camera, Thomson parabola spectrographs, a Rutherford magnetic spectrograph, plasma visible spectroscopy, and several nuclear activation diagnostics. These components, when taken together, enable a rather thorough description of the 5-MV, 10-TW ion beams presently being produced. A unique feature of these diagnostics is that they are capable of operating in hard (several MeV) x-ray bremsstrahlung backgrounds of some 10/sup 9/--10/sup 10/ rad/s. Details of each diagnostic, its integration, data reduction procedures, and recent PBFA-II data will be discussed.
Research Organization:
Sandia National Laboratories, Albuquerque, New Mexico 87185
OSTI ID:
6949782
Journal Information:
Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 59:8; ISSN RSINA
Country of Publication:
United States
Language:
English