Backflush column removes GC sample water

Journal Article · · Hydrocarbon Processing; (USA)
OSTI ID:6929259
 [1]
  1. Shell Refinery, Pernis (NL)

Plots columns coated with Al{sub 2}O{sub 3}/KCl are a good tool for separating light hydrocarbons (C{sub 1} to C{sub 10}). A disadvantage of these types of columns, however, is their sensitivity for water. Water deactivates the Al{sub 2}O{sub 3} which results in shortened retention times. This is a reversible process resulting in a continuous shift of retention times, which makes these types of columns unattractive for us in process GCs. This article deals with a column configuration to evade this problem.

OSTI ID:
6929259
Journal Information:
Hydrocarbon Processing; (USA), Journal Name: Hydrocarbon Processing; (USA) Vol. 69:3; ISSN HYPRA; ISSN 0018-8190
Country of Publication:
United States
Language:
English