A 2-dimensional detector decoding study on BGO arrays with quadrant sharing photomultipliers
Conference
·
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:6838599
- Univ. of Texas, Houston, TX (United States). MD Anderson Cancer Center
This is a theoretical estimation and experimental study of the positron camera detector design using quadrant sharing of phototubes and optically cross-coupled BGO crystals which can be coupled to a photomultiplier and be decoded with this design, based on the number of photoelectrons generated and the optical efficiency of the crystal-phototube combination. From photoelectron statistics, a 10 x 10 BGO array can be coupled to a single square phototube and have each crystal clearly decoded. Simulation of 1-dimensional and 2-dimensional statistical photoelectron distribution verifies the estimations. The simulation also yields the optimal distribution of light to the decoding phototubes. This theoretically optimal light distribution is used to guide the development of an 8 x 8 array of 3.2 x 3.2 mm BGO crystals using Hamamatsu R-2497 phototubes. All 64 crystals in the array can be clearly identified. The reconstructed image resolution is further estimated for the 8 x 8 and 10 x 10 arrays. These resolution results are also compared to discrete detectors which do not use analog decoding. The comparison shows that the analog decoding process degrades the image resolution by 7 to 9%.
- OSTI ID:
- 6838599
- Report Number(s):
- CONF-931051--
- Conference Information:
- Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Journal Volume: 41:4PT1
- Country of Publication:
- United States
- Language:
- English
Similar Records
An analog decoding BGO block detector using circular photomultipliers
Collection of scintillation light from small BGO crystals
Initial characterization of a bgo-Photodiode detector for high resolution positron emission tomography
Journal Article
·
Tue Aug 01 00:00:00 EDT 1995
· IEEE Transactions on Nuclear Science
·
OSTI ID:148054
Collection of scintillation light from small BGO crystals
Journal Article
·
Tue Aug 01 00:00:00 EDT 1995
· IEEE Transactions on Nuclear Science
·
OSTI ID:148048
Initial characterization of a bgo-Photodiode detector for high resolution positron emission tomography
Conference
·
Tue Jan 31 23:00:00 EST 1984
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:6850524
Related Subjects
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
BGO DETECTORS
CAMERAS
COMPUTERIZED TOMOGRAPHY
DESIGN
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
GAMMA CAMERAS
MEASURING INSTRUMENTS
OPTIMIZATION
POSITRON CAMERAS
POSITRON COMPUTED TOMOGRAPHY
RADIATION DETECTORS
RESOLUTION
SCINTILLATION COUNTERS
SOLID SCINTILLATION DETECTORS
TOMOGRAPHY
62 RADIOLOGY AND NUCLEAR MEDICINE
BGO DETECTORS
CAMERAS
COMPUTERIZED TOMOGRAPHY
DESIGN
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
GAMMA CAMERAS
MEASURING INSTRUMENTS
OPTIMIZATION
POSITRON CAMERAS
POSITRON COMPUTED TOMOGRAPHY
RADIATION DETECTORS
RESOLUTION
SCINTILLATION COUNTERS
SOLID SCINTILLATION DETECTORS
TOMOGRAPHY