A new compact position-sensitive PMT for scintillation detectors
Journal Article
·
· IEEE Transactions on Nuclear Science
- Hamamatsu Photonics K.K., Toyooka (Japan). Electron Tube Center
- Hamamatsu Photonics K.K., Hirakuchi (Japan). Central Research Lab.
- Zhejiang Univ. (China). State Key Lab. of Modern Optical Instrumentation
A new compact position-sensitive photomultiplier tube (PS-PMT), Hamamatsu R7600-C12, has been developed for scintillation detectors. The PS-PMT has 11 stages of metal channel dynodes and 6(X) + 6(Y) crossed plate anodes in a 25.7 mm square x 20 mm high metal can package, where the photo-sensitive area is 22 mm square. The performance of the PS-PMT was evaluated in terms of applicability to radiation imaging systems. In comparison with the former type of PS-PMT (Hamamatsu R5900-C8), the new PS-PMT provides smaller light spread and better position response. Also, by removing the flange at the bottom of the PS-PMT, the ratio of the effective area to the outward area is increased. The spatial resolution capability was demonstrated by imaging a stratified LSO array having an element of 1.8 mm x 1.8 mm x 10 mm. Each crystal element is clearly identified on the image map with 662 keV uniform gamma-ray irradiation. This paper describes the characteristics of the new PS-PMT and the experimental results for a gamma-ray imaging detector.
- OSTI ID:
- 679583
- Report Number(s):
- CONF-981110--
- Journal Information:
- IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 3Pt1 Vol. 46; ISSN 0018-9499; ISSN IETNAE
- Country of Publication:
- United States
- Language:
- English
Similar Records
Characterization of a new multianode PMT for low-level optical fiber readout
Multi-PSPMT scintillation camera
A compact position-sensitive detector for PET
Conference
·
Mon Dec 30 23:00:00 EST 1996
·
OSTI ID:513060
Multi-PSPMT scintillation camera
Journal Article
·
Tue Jun 01 00:00:00 EDT 1999
· IEEE Transactions on Nuclear Science
·
OSTI ID:684505
A compact position-sensitive detector for PET
Journal Article
·
Tue Aug 01 00:00:00 EDT 1995
· IEEE Transactions on Nuclear Science
·
OSTI ID:148053