Title: Computed tomographic scanning in children: comparison of radiation dose and resolving power of commercial CT scanners

Conference · · Am. J. Roentgenol.; (United States)
OSTI ID:6788180

Surface and internal radiation doses for abdominal computed tomography (CT) of children were determined using child-sized phantoms and seven models of CT body scanners. Resolving power of each scanner was determined simultaneously with the radiation dose determination. The average surface skin dose for a complete CT body examination ranged from 0.39 to 5.60 rad, varying with patient size and model of CT scanner employed. A high contrast (12%) resolving power phantom of water-filled holes in acrylic showed a range of 1.75 to 2.25 mm.

Research Organization:
Univ. of California, San Francisco
OSTI ID:
6788180
Journal Information:
Am. J. Roentgenol.; (United States), Vol. 131:1; Conference: Annual meeting of the Society for Pediatric Radiology, Boston, MA, USA, Sep 1977
Country of Publication:
United States
Language:
English