Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Dosimetry of a dual photon energy linear accelerator

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.596083· OSTI ID:6746889
We describe the dosimetric characteristics of a newly introduced dual photon energy linear accelerator, the Varian Clinac 1800. Depth doses are compared with other accelerators of the same nominal accelerating potentials (6 and 10 MV). Field flatness at d/sub max/ and at 10 cm depth, depth of d/sub max/, wedge characteristics, output factors, and doses in the buildup region are presented.
Research Organization:
Department of Radiation Physics, City of Hope National Medical Center, Duarte, California 91010
OSTI ID:
6746889
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 14:2; ISSN MPHYA
Country of Publication:
United States
Language:
English