Calorimetric and ionimetric dosimetry intercomparisons I: U. S. neutron radiotherapy centers

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.594905· OSTI ID:6744949
In the U.S. neutron radiotherapy trial centers, absorbed dose is routinely measured using commercially available A-150 tissue equivalent (TE) plastic ionization chambers. The collecting volumes of these chambers are filled with either methane-based tissue equivalent gas or air. Absorbed dose in A-150 plastic, determined with these ionization chambers, was compared to that measured by an A-150 plastic calorimeter in an A-150 plastic phantom. These comparisons have yielded the following information: (1) Agreement of the total absorbed dose measured using the ionization chambers was within 2.5% of the calorimeter at all the centers visited to date. (2) For all the neutron fields measured, the product of the stopping power ratio (s/sub w/,g)/sub N/' between the A-150 plastic chamber wall and TE gas, and the average energy expended in the gas per ion pair formed, W-bar/sub N//e, was computed assuming Bragg--Gray theory and found to be 31.0 +- 0.7 J/C. (3) The displacement correction factor employed to normalize measurements at a depth in a phantom using the type IC-17 ionization chamber was verified to be approximately 0.97 +- 0.01.
Research Organization:
Radiation Biophysics Laboratory, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, New York 10021
OSTI ID:
6744949
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 8:1; ISSN MPHYA
Country of Publication:
United States
Language:
English