Performance of a BaF[sub 2]-TMAE prototype detector for use in PET
Conference
·
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:6698975
- Royal Marsden Hospital, Belmont (United Kingdom). Dept. of Physics
- Rutherford Appleton Lab., Chilton (United Kingdom)
A BaF[sub 2]-TMAE detector has been developed for use in positron emission tomography (PET). The detector achieves an efficiency of 23.5%, a time resolution of 3.5ns FWHM, and a spatial resolution of 6.0mm FWHM with no change in the normal operating parameters. The only deterioration observed in a typical working day is a small increase in the FWTM of the time resolution. This paper describes how this level of performance was achieved and also how these results have been used as the basis for the design of a large area positron camera.
- OSTI ID:
- 6698975
- Report Number(s):
- CONF-931051--
- Conference Information:
- Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Journal Volume: 41:6Pt2
- Country of Publication:
- United States
- Language:
- English
Similar Records
A prototype BaF sub 2 /TMAE low pressure multiwire proportional chamber for PET
PETRRA: Preliminary experimental results from the first full size detector and dead time simulation of the count rate performance of a unique whole body PET camera
Test of a BaF/sub 2/-TMAE detector for positron-emission tomography
Journal Article
·
Sun Mar 31 23:00:00 EST 1991
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
·
OSTI ID:5012756
PETRRA: Preliminary experimental results from the first full size detector and dead time simulation of the count rate performance of a unique whole body PET camera
Journal Article
·
Tue Aug 01 00:00:00 EDT 1995
· IEEE Transactions on Nuclear Science
·
OSTI ID:148044
Test of a BaF/sub 2/-TMAE detector for positron-emission tomography
Conference
·
Sat Jan 31 19:00:00 EST 1987
· IEEE Trans. Nucl. Sci.; (United States)
·
OSTI ID:6726259
Related Subjects
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
CAMERAS
CHARGED PARTICLE DETECTION
COMPUTERIZED TOMOGRAPHY
DETECTION
DIAGNOSTIC TECHNIQUES
EFFICIENCY
EMISSION COMPUTED TOMOGRAPHY
GAMMA CAMERAS
MEASURING INSTRUMENTS
PERFORMANCE
POSITRON CAMERAS
POSITRON COMPUTED TOMOGRAPHY
POSITRON DETECTION
RADIATION DETECTION
RADIATION DETECTORS
TOMOGRAPHY
62 RADIOLOGY AND NUCLEAR MEDICINE
CAMERAS
CHARGED PARTICLE DETECTION
COMPUTERIZED TOMOGRAPHY
DETECTION
DIAGNOSTIC TECHNIQUES
EFFICIENCY
EMISSION COMPUTED TOMOGRAPHY
GAMMA CAMERAS
MEASURING INSTRUMENTS
PERFORMANCE
POSITRON CAMERAS
POSITRON COMPUTED TOMOGRAPHY
POSITRON DETECTION
RADIATION DETECTION
RADIATION DETECTORS
TOMOGRAPHY