A transportable x-ray fluorescence analysis system for assay of plutonium and uranium solutions in gloveboxes
- Lawrence Livermore National Lab., CA (USA)
- Commission of the European Communities, Luxembourg (Luxembourg). Safeguards Directorate
- Commission of the European Communities, Karlsruhe (Germany, F.R.). European Inst. for Transuranium Elements
It is becoming increasingly more difficult and expensive for safeguards authorities to transport solution samples from inspected facilities to laboratories for analysis. We have developed a transportable x-ray fluorescence analysis (XRFA) instrument that can assay plutonium and uranium solutions in glove boxes without any modifications to the glove box. This instrument will reduce the need to transport plutonium solution samples for analysis, thus saving time and reducing costs in inspection efforts. The system utilizes a portable high-purity germanium detector, transportable pulse-processing electronics, and a portable IBM-compatible computer. The measurement head of our instrument is placed inside a glove-box glove. Inside the glove box a hollow cylinder is held in place by the measurement head with the glove partially surrounding it, then solutions contained in 15-mL polystyrene test tubes are inserted into the cylinder for assay by XRFA. The radioisotope {sup 57}Co is used to excite the uranium and plutonium K-series x rays, and a combined {sup 57}Co and{sup 153}Gd source is used for transmission measurements to correct the observed x-ray intensity for sample self-attenuation effects. We have tested and evaluated the instrument at the European Institute for Transuranium Elements and demonstrated that 1% accuracies are achievable using eight uranium and seven plutonium reference solutions ranging in concentration from 50 to 411 g/L and 50 to 303 g/L, respectively. 3 refs., 2 figs., 2 tabs.
- Research Organization:
- Lawrence Livermore National Lab., CA (USA)
- Sponsoring Organization:
- DOE/DP
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 6683326
- Report Number(s):
- UCRL-102705; CONF-9007106-55; ON: DE90015206
- Resource Relation:
- Conference: Institute of nuclear materials management conference, Los Angeles, CA (USA), 15-18 Jul 1990
- Country of Publication:
- United States
- Language:
- English
Similar Records
Semiannual Progress Report for the Period July 1960 through December 1960
Mechanical Design of Hybrid Densitometer for Laboratory Applications
Related Subjects
99 GENERAL AND MISCELLANEOUS//MATHEMATICS, COMPUTING, AND INFORMATION SCIENCE
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
37 INORGANIC
ORGANIC
PHYSICAL AND ANALYTICAL CHEMISTRY
NUCLEAR MATERIALS MANAGEMENT
SAFEGUARDS
RADIOISOTOPES
QUALITATIVE CHEMICAL ANALYSIS
X-RAY FLUORESCENCE ANALYZERS
PERFORMANCE TESTING
ACCURACY
COBALT 57
COMPARATIVE EVALUATIONS
CONCENTRATION RATIO
ECONOMICS
GADOLINIUM 153
GLOVEBOXES
PLUTONIUM
PORTABLE EQUIPMENT
SAMPLING
URANIUM
X-RAY FLUORESCENCE ANALYSIS
ACTINIDES
BETA DECAY RADIOISOTOPES
CHEMICAL ANALYSIS
COBALT ISOTOPES
DAYS LIVING RADIOISOTOPES
ELECTRON CAPTURE RADIOISOTOPES
ELEMENTS
EQUIPMENT
EVEN-ODD NUCLEI
GADOLINIUM ISOTOPES
INTERMEDIATE MASS NUCLEI
ISOTOPES
LABORATORY EQUIPMENT
MANAGEMENT
METALS
NONDESTRUCTIVE ANALYSIS
NUCLEI
ODD-EVEN NUCLEI
RARE EARTH ISOTOPES
RARE EARTH NUCLEI
TESTING
TRANSURANIUM ELEMENTS
X-RAY EMISSION ANALYSIS
055001* - Nuclear Fuels- Safeguards
Inspection
& Accountability- Technical Aspects
990200 - Mathematics & Computers
440101 - Radiation Instrumentation- General Detectors or Monitors & Radiometric Instruments
400101 - Activation
Nuclear Reaction
Radiometric & Radiochemical Procedures