Title: At-line benzene monitor for measuring benzene in precipitate hydrolysis aqueous

Technical Report ·
DOI:https://doi.org/10.2172/6678206· OSTI ID:6678206

A highly accurate and repeatable at-line benzene monitor (ALBM) has been developed to measure the benzene concentration in precipitate hydrolysis aqueous (PHA) in the DWPF. This analyzer was conceived and jointly developed within SRTC by the Analytical Development and the Defense Waste Process Technology Sections with extensive support from the Applied Statistics Group and the TNX Operations Section. It is recommended that an ALBM specifically adapted to DWPF analytical requirements be used to measure benzene in PHA; calibrations be performed using a 10% methanol solution matrix (for standard stability); and based on experience gained in development at TNX, the services of ADS and ASG be employed to both adapt the ALBM to DWPF requirements and develop statistical control procedures.

Research Organization:
Savannah River Site (SRS), Aiken, SC (United States)
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
AC09-89SR18035
OSTI ID:
6678206
Report Number(s):
WSRC-RP-92-1215; ON: DE93008936
Country of Publication:
United States
Language:
English