Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Nationwide evaluation of /sup 60/Co teletherapy calibrations

Journal Article · · Med. Phys.; (United States)
OSTI ID:6667978
As a result of a reported incident which resulted in the unintended overexposure of several hundred patients, teletherapy installations throughout the United States were evaluated to verify the accuracy of teletherapy equipment calibrations. Selected licensees received TLDs for exposure under specified conditions. Site visits were made to each licensee showing an apparent discrepancy greater than +- 5% and intercomparison measurements made. This paper describes results of a sampling of 373 teletherapy licensees and the success of site visits in reducing final discrepancies to less than +- 5%. TLD and site visit measurement techniques, results, comparison with data from other studies, and possible sources of uncertainty in the management of a large-scale screening program are discussed.
Research Organization:
United States Nuclear Regulatory Commission, 799 Roosevelt Road, Glen Ellyn, Illinois 60137.
OSTI ID:
6667978
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 5:5; ISSN MPHYA
Country of Publication:
United States
Language:
English