Measurement of neutron dose and spectra at light-water reactors

Conference ·
OSTI ID:6663516
In choosing one dosimeter over another, both sensitivity and precision must be addressed. It has been determined that dosimeters employing NTA film lack adequate sensitivity for use inside containment of nuclear plants. From this study, it is apparent that CR-39 and polycarbonate track etch films used without radiators are also inadequate. The rest of the dosimeters tested displayed adequate sensitivity. The two general types of dosimeters which comprise that group are: (1) TLDs and (2) the polycarbonate track etch which was used in conjunction with (n, ..cap alpha..) radiators. The precision of the TLD dosimeter depends on the calibration technique and response correction technique. TLD dosimeters calibrated to D/sub 2/O-moderated californium-252 or corrected based on 9- to 3-in. sphere response rates seemed to function best inside reactor containment. Polycarbonate track etch film with (n, ..cap alpha..) radiators was found to be sensitive enough for use inside reactor containment; however, the standard deviation for the results was roughly twice the value of the most precise TLD albedo dosimeters. Also, several of the polycarbonate dosimeters saturated, rendering any evaluation impossible.
Research Organization:
Pacific Northwest Lab., Richland, WA (USA)
DOE Contract Number:
AC06-76RL01830
OSTI ID:
6663516
Report Number(s):
PNL-SA-10728; CONF-821037-40; ON: DE83002691