Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Gas chromatographic determination of gases formed in catalytic reduction of nitric oxide

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00229a018· OSTI ID:6582957

The construction and performance are described of a gas chromatographic system for the analysis of gas mixtures formed during the catalytic reduction of NO with CO, H/sub 2/, and NH/sub 3/. Reactants and products are separated by using three packed columns contained within two isothermally operated gas chromatographs. The system operation, data acquisition, and data reduction are accomplished by means of a microprocessor. Quantitative analysis of NO, NH/sub 3/, N/sub 2/, N/sub 2/O, CO, and CO/sub 2/ can be obtained down to levels in the range of 10 to 35 ppM, and H/sub 2/ can be analyzed down to 400 ppM. Water present in the sample is separated from the other components but is not quantified.

Research Organization:
Univ. of California, Berkeley
OSTI ID:
6582957
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 53:6; ISSN ANCHA
Country of Publication:
United States
Language:
English