skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Performance testing of Eberline Alpha 6 and Alpha 6A continuous air monitors

Technical Report ·
DOI:https://doi.org/10.2172/658287· OSTI ID:658287

Eberline Alpha 6 and Alpha 6A continuous air monitors (CAMs) were tested against the performance criteria of the International Electrotechnical Commission standard 761-6, ``Equipment for Continuously Monitoring Radioactivity in Gaseous Effluents, Part 6: Specific Requirements for Transuranic Aerosol Effluent Monitors``, and against ANSI N42.17B, ``Performance Specification Health Physics Instrumentation--Occupational Airborne Radioactivity Monitoring Instrumentation``. The performance criteria require the CAM`s response to a radioactive source to remain within a tolerance while the CAM is exposed to an external influence such as temperature, electromagnetic fields, or ionizing radiations. The CAMs complied within specified tolerances with a majority of the performance specifications. The most significant problems with CAM performance were noted during exposures to external nonionizing radiation fields (radio frequency fields). At numerous frequencies, the CAMs did not respond to radioactive material in the filter holder. At other frequencies and in some orientations, the CAMs overresponded by orders of magnitude. In addition to sensitivity to external nonionizing radiation fields, the CAMs exhibited sensitivity to electrostatic discharges.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE Assistant Secretary for Environment, Safety, and Health, Washington, DC (United States)
DOE Contract Number:
AC06-76RL01830
OSTI ID:
658287
Report Number(s):
PNNL-11527; ON: DE98053052; TRN: 98:011214
Resource Relation:
Other Information: PBD: Jan 1998
Country of Publication:
United States
Language:
English