Apparatus for measurement of surface resistance versus rf magnetic field of high- Tc superconductors

Journal Article · · Review of Scientific Instruments; (USA)
DOI:https://doi.org/10.1063/1.1141390· OSTI ID:6544378
; ;  [1]
  1. Argonne National Laboratory, Engineering Physics Division, 9700 South Cass Avenue, Argonne, IL (USA)

A coaxial niobium quarter-wave resonant cavity has been fabricated and used for measurement of the surface resistance of disk-shaped samples of high-{Tc} superconductors at 820 MHz and 4.2 K, and at surface rf magnetic fields ranging to 300 G. The cavity can accommodate samples as large as 36 mm in diameter. A measurement procedure has been developed with which rf surface resistances less than 20 {mu}{Omega} can be resolved for a 24-mm-diam sample at 820 MHz. This cavity is useful in determining the suitability of high-{ital T}{sub {ital c}} superconductors for high-power rf applications.