Surface dose for megavoltage photon beams outside the treatment field
Journal Article
·
· Med. Phys.; (United States)
Measurements made on photon beams from four different radiotherapy machines have demonstrated that skin dose several centimeters outside the boundary of a treatment field may be as much as 20% of the central axis maximum dose. This surface dose has been measured for an AECL Theratron 80, Siemens Mevatron VI, Varian Clinac 20, and CGR Sagittaire for distances up to 12 cm outside the field boundary and for depths up to the depth of maximum central axis dose. This dose has also been measured as a function of field size and of source-to-skin distance. For the lower energy photon beams, this radiation is significantly attenuated in the first 2-3 mm of tissue, while for higher energy beams, a buildup phenomenon with a dmax of 2-3 mm is observed. The magnitude of this radiation is approximately linearly dependent upon field dimension for all energies.
- Research Organization:
- Univ. of Kansas Medical Center, Kansas City
- OSTI ID:
- 6537324
- Journal Information:
- Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 10:6; ISSN MPHYA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Surface dose for megavoltage photon beams outside the treatment field
Surface dose from megavoltage therapy machines
Dosimetry of the Siemens Mevatron 67 linear accelerator
Journal Article
·
Mon Oct 31 23:00:00 EST 1983
· Med. Phys.; (United States)
·
OSTI ID:5516081
Surface dose from megavoltage therapy machines
Journal Article
·
Sun Nov 30 23:00:00 EST 1975
· Radiology, v. 117, no. 3, pp. 705-708
·
OSTI ID:4167930
Dosimetry of the Siemens Mevatron 67 linear accelerator
Journal Article
·
Mon Aug 01 00:00:00 EDT 1983
· Int. J. Radiat. Oncol., Biol. Phys.; (United States)
·
OSTI ID:5531261